IO0062 - VENTICELL 55 R
- - Accurate and quick laboratory oven
- - 55-liter chamber volume
- - Working temperature of from +10°C over ambient temperature up to 250°C300 °C
- - Patented system of forced air flow in the chamber ensures a homogenous temperature profile for all processes of drying, heating and sterilizing materials in laboratories
- - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
- - Greater speed and accuracy of all tempering procedures ensures cost-saving operation; suitable especially for materials of high humidity, for demanding and accurate tests and processes
- - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
- - Wide range of optional equipment and accessories
- - Output for a printer or PC connection
- - Validation possibility (IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
The values may differ depending on specific charge and media parameters
- volume 55 l
- depth 390 mm
- width 400 mm
- height 350 mm
Tray
- standard equipment 2 pcs
- storage area (w × d) 380×335 mm
- racks 4 max. number
- distance between guide rails 70 mm
Maximal weight of the load *)
- inside the oven 50 kg/oven
- per tray 20 kg/tray
Number of doors
- 1 pcs
External dimensions (including door and handle)
- depth 640 mm
- diameter of the air branch outer/inner 52/49 mm
- width 620 mm
- height, incl. Feet (N) and Caster (K) 680 (N) mm
Package dimensions (three layers carton)
- depth 730 mm
- width 700 mm
- depth 880 mm
Weight
- gross 66 kg
- net 55 kg
Working temperature **) (beginning of the regulation) - VENTICELL / VENTICELL plus
- from 10 °C above ambient temp. to 250 / 300 °C
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 °C
- time variation cca (±) 0,3 °C
- space deviation (±) % cca (±)% of the reached temperature 1 %
Time required to reach 250 °C with closed air flap and voltage 230 V
- 49 min
Heat emission at 250 °C
- 590 W
Air exchange speed at 250 °C
- 45 cca/hour
Electricity - mains 50/60 Hz
- max. power input 1,3 kW
- nominal voltage 230 V
- current 5,6 A
- stand by 5 W
*) Approx. 50 % of the tray area can be filled while maintaining uniform air circulation inside the chamber.
**) Standard type is up to 250 °C, optional type is up to 300 °C.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
The values may differ depending on specific charge and media parameters