II0033 - INCUCELL 222 - Standard
- - Natural, precise – laboratory incubator
- - 222-liter chamber volume
- - Working temperature of from +5°C above ambient up to 99,9°C
- - Noiseless operation and fine air flow in the chamber
- - High operation comfort, precise temperature regulation and short times of temperature recovery in the chamber after the door opening
- - High-quality and safe drying, heating and tempering of common laboratory materials in laboratories and industry; suitable for safe processing of microbiological cultures
- - Standard model with a Fuzzy-Logic microprocessor control system; control elements situated on a foil keyboard, process information indicated on a LED display
- - Wide range of optional equipment and accessories
- - Output for a printer or PC connection
- - Validation possibility (IQ , OQ)
Technical data
Interior of stainless steel material No. 1.4301
- volume 222 l
- depth 520 mm
- width 540 mm
- height 760 mm
Tray
- standard equipment 2 pcs
- storage area (w × d) 520×485 mm
- racks 10 max. number
- distance between guide rails 70 mm
Maximal weight of the load*)
- inside the oven 70 kg/oven
- per tray 30 kg/tray
Door
- 1 pcs
External dimensions (including door and handle)
- depth 790 mm
- diameter of the air branch outer / inner 52/49 mm
- width 760 mm
- height, incl. Feet (N) and Caster (K) 1090 (N)
Package dimensions (three layers carton)
- depth 860 mm
- width 830 mm
- depth 1280 mm
Weight
- gross 116 kg
- nett 100 kg
Working temperature (beginning of the regulation)
- from 5 °C above ambient temp. to: 99,9 °C
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
- time variation cca (±) 0,2 °C
- space deviation cca (±) 1 °C
Time required to reach 37 °C with closed air flap and voltage 230 V
- 79 min
Heat emission at 37 °C
- 45 W
Air exchange speed at 37 °C
- 5 per hour
Electricity - mains 50/60 Hz
- max. power 0,5 kW
- nominal voltage 230 V
- current 2 A
- power by stand by 5 W
*) Approx. 50 % of the tray area can be filled while allowing uniform air circulation inside the chamber.
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.